Model | CHW700 | CHS700 |
|---|
Performance
|
Operating Water | For city water circulation | For pure water circulation |
Circulation Type | Closed system circulation by pump |
Operating Temperature Range | 10癈 to Room temp. |
Tem. Adjustment Accuracy | ?.5 to 2.5癈 |
Cooling Capacity | Approx. 1000W (860kcal) at Liquid temp. 20癈 at Room temp. 20癈 |
Components
|
Temp. Control System | Refrigerator On/Off control |
Temp. Sensor | T-thermocouple |
Temp. Setting Display | Digital |
Refrigerator 旵oolant | Air cooling 675W R407C |
Circulation Pump | Magnet pump 65W x 2pcs. |
Pump Max. Flow Rate | 22/25L/min. (50/60Hz) |
Pump Max. Lift | 10/13.5m (50/60Hz) |
Surrounding Operation Temp. Range | 5 to 30癈 |
Circulation Nozzle | Rc3/8 Female screw |
Safety Device
|
Safety Countermeasures | Self diagnosis functions Electric leakage breaker with over current protection Refrigerator over load relay Refrigerator pressure detector Refrigerator protection抯 delay timer function Float switch Circulation pump protection by-pass |
Other Function | Drain cock Dust prevention filter Refrigerator pressure indicator Automatic stop operation Automatic start operation Key lock function Calibration off set function Breakout protection function Temp. output terminal (4-20mA). |
Accessories | Drain hose 0.5m x 1pc. |
Standard
|
Water Bath Dimension | ID300xH236mm |
Water Bath Capacity | 14L |
Cooling Coil | Copper-nickel plated | SUS304 |
External Dimension | W380 x D460 x H1050 mm |
Power Source | AC100V 50/60Hz 12A |
Weight | Approx. 55kg |